Skip to main content

Read the gas concentration

What this example shows​

The electrochemical-bridgeBoard.py example starts the board at its default address, 0x30, sets up the analog front end for one gas cell, and prints the gas concentration in parts per million (PPM) every 2.5 seconds.

Behind address 0x30 is the board's ATtiny404. It configures the LMP91000 and reads the ADS1115 on the board's own internal bus, so your script only ever talks to one I2C device.

The example code​

The example uses the ozone preset, SENSOR_O3, and opens I2C bus 0 with SCL on pin 22 and SDA on pin 21. Change the pins to match your controller, and import and pass the preset for your cell: SENSOR_CO, SENSOR_NO2, SENSOR_SO2, SENSOR_NO, SENSOR_H2S, SENSOR_NH3 or SENSOR_CL2. The example's comments call the address setting "jumpers on the back of the board"; on this board it is the ADDRESS switch on the top. If you set it to another address, change adc_addr to match. The comments also say LMPEN is tied to GND; on this board it goes to the header and to the ATtiny404, and the library does not drive it. The comment that mentions "legacy boards with 0x49/0x4A/0x4B" means boards read at those addresses, where the library talks to an ADS1115 directly instead of to the bridge. This board answers at 0x30 to 0x37.

# FILE: electrochemical-bridgeBoard.py
# AUTHOR: Fran Fodor @ Soldered
# BRIEF: Single sensor PPM reading example for the ATtiny404 bridge board revision,
# which allows daisy-chaining multiple boards on one I2C bus.
# LAST UPDATED: 2026-07-23

# Connecting diagram:
#
# Qwiic------------------------>Qwiic
#
# Note: on this board revision, LMPEN is hardwired to GND, so there's no
# config_pin to set - it's ignored whenever adc_addr falls in the bridge
# range (0x30-0x37). Set the desired address using the jumpers on the back
# of the board before powering it on.

from machine import I2C, Pin
import time
from ElectrochemicalGasSensor import ElectrochemicalGasSensor, SENSOR_O3

# Initialize I2C
i2c = I2C(0, scl=Pin(22), sda=Pin(21))

# (address = 0x30 + jumper offset, giving a range of 0x30-0x37).
# Any additional boards on the same bus just need a different jumper
# setting, same as twoSensors.py does for legacy boards with 0x49/0x4A/0x4B.
sensor = ElectrochemicalGasSensor(SENSOR_O3, i2c, adc_addr=0x30)

# -------------------------------------------------------------------------
# Setup
# -------------------------------------------------------------------------
print("Electrochemical Gas Sensor - ATtiny Bridge Board Example")

if not sensor.begin():
print("ERROR: Can't init the sensor! Check connections and jumper address!")
while True:
time.sleep_ms(100)

print("Sensor initialized successfully!")

# -------------------------------------------------------------------------
# Main loop
# -------------------------------------------------------------------------
while True:
reading = sensor.getPPM()

# If PPB is more relevant for your sensor, use:
# reading = sensor.getPPB()

print("Sensor reading: {:.5f} PPM".format(reading))

time.sleep_ms(2500)

🔗Open the complete example

Examples/electrochemical-bridgeBoard.py at commit 623e1e6

→

Expected result, in the shell:

ShellMicroPython
Electrochemical Gas Sensor - ATtiny Bridge Board Example
Sensor initialized successfully!

Then a line every 2.5 seconds: Sensor reading: , the concentration with five decimals, and PPM. If the board does not answer, the script prints ERROR: Can't init the sensor! Check connections and jumper address! and waits.

Functions used​

I2C and Pin come from MicroPython's own machine module, and time.sleep_ms() from its time module. The functions below are the library's.

ElectrochemicalGasSensor(sensor_type, i2c=None, adc_addr=0x49, config_pin=None)returns ElectrochemicalGasSensor

Creates the sensor object for one board. The library reads adc_addr to decide how to talk to it: for an address from 0x30 to 0x37 it uses the protocol of the ATtiny404 bridge; for any other address it talks to an ADS1115 at that address and to the LMP91000 at 0x48 itself. The example passes 0x30.

ReturnsThe sensor object.

Parameters

NameTypeDescription
sensor_typeSensorConfigThe preset for your cell: SENSOR_CO, SENSOR_NO2, SENSOR_SO2, SENSOR_O3, SENSOR_NO, SENSOR_H2S, SENSOR_NH3, SENSOR_CL2, or your own SensorConfig.
i2cmachine.I2CThe I2C bus the board is on. When left out, the library opens I2C bus 0 with SCL on pin 22 and SDA on pin 21 on an ESP32 port, or I2C with SCL on pin 5 and SDA on pin 4 on an ESP8266 port, and raises an exception on any other board.
adc_addrintThe board's I2C address, 0x30 to 0x37 as set by the address switch. 0x30 out of the box.
config_pinintA GPIO number for the LMP91000 enable line. The library holds it high and pulls it low only while it writes the LMP91000's registers itself; for an address from 0x30 to 0x37 it is not used, so the example leaves it out. None by default.
begin()returns bool

Checks that the bridge answers, sends the ADC gain from the preset, and configures the LMP91000 with the preset's gain, load, reference, bias and mode. Call it once before reading.

ReturnsTrue when the bridge answered and both configurations succeeded, False otherwise. Each step gives up after 500 ms without an answer.

getPPM()returns float

Takes one ADC reading through the bridge and converts it to a concentration with the preset: it removes the internal zero offset, adds the zero calibration, divides by the amplifier gain and by the cell's sensitivity in nA per PPM. A negative result is returned as 0.

ReturnsThe gas concentration in PPM.

getPPB()returns float

The same reading as getPPM(), in parts per billion.

ReturnsThe gas concentration in PPB, getPPM() x 1000.

Putting it together​

Create one ElectrochemicalGasSensor per board, with the preset for its cell, the I2C bus and its address. Call begin() once and stop if it returns False. Then call getPPM() or getPPB() as often as you need a reading.

To run more than one board on the same bus, give each its own address with the address switch (0x30 to 0x37) and create one object per address on the same i2c bus.