Calibrate the zero point

What this example shows​

The calibrateSensor example applies a zero-point offset and prints the gas concentration in PPM every 2.5 seconds. It creates the sensor with address 0x49 and configuration pin 32.

customCalibration is a voltage the library adds to the ADC voltage before it converts it. The example's comment gives the method: in an environment with none of the target gas, adjust the value until the voltage after calibration is roughly 0.

To see that voltage, the example asks you to uncomment one line in the library's header file, Electrochemical-Gas-Sensor-SOLDERED.h:

//#define ELECTROCHEMICAL_SENSOR_DEBUG

With it defined, every getPPM() call also prints the raw voltage measurement, the voltage without the reference value, and the voltage after calibration.

The example is written with SENSOR_NH3, the library's ammonia preset, so its PPM values are not H2S values. Do not switch it to the library's SENSOR_H2S preset unchanged: that preset is written for the SGX-4H2S-100 cell, with a sensitivity of 1200 nA per ppm, while this board carries the SGX-4H2S-300 cell, whose datasheet gives an output signal of 460 ± 140 nA per ppm. With SENSOR_H2S the reading comes out about 2.6 times too low. Pass a custom sensor type with the SENSOR_H2S settings and 460 nA per ppm instead, as Quick Start shows.

The example code​

Select Board
calibrateSensor.ino
// Include the required library
#include "Electrochemical-Gas-Sensor-SOLDERED.h"

// This is the custom value we set for the calibration voltage
// This value is added to the voltage read from the ADC
// In an environment with 0 of the target gas, the voltage after calibration should be
// roughly 0, adjust customCalibration to get it to that value.
const double customCalibration = 0.05;


// Create the sensor object with the according type
ElectrochemicalGasSensor sensor(SENSOR_NH3, 0x49, 32);



void setup()
{
Serial.begin(115200); // For debugging

// Initialize the breakout
if (!sensor.begin())
{
// Can't init? Notify the user and go to infinite loop
Serial.println("ERROR: Can't init the sensor! Check connections!");
while (true)
delay(100);
}

//Set the custom calibration value
sensor.setCustomZeroCalibration(customCalibration);

Serial.println("Sensor initialized successfully!");
}

void loop()
{
// Make the reading
double reading = sensor.getPPM();
// This function will also print debug messages if the #define in
// the header file of the library is uncommented
// It will print:
// - The raw voltage measurement
// - Voltage without reference value
// - Voltage after calibration
// These values can help you adjust the internal zero calibration



// Print the reading with 5 digits of precision
Serial.print("Sensor reading: ");
Serial.print(reading, 5);
Serial.println(" PPM");

// Wait a bit before reading again
delay(2500);
}

Expected result:

Serial MonitorAutoscrollShow timestampNewline115200 baud
Message (Enter to send message to the board)
Sensor initialized successfully!

Then a line starting with Sensor reading: follows every 2.5 seconds, with the concentration in PPM to five decimals. The library never reports a value below zero: it returns 0 instead, so 0.00000 is a normal reading. With ELECTROCHEMICAL_SENSOR_DEBUG defined, each reading is preceded by the lines Electrochemical gas sensor readings:, Raw voltage measurement:, Voltage without reference value: and Voltage after calibration:, each voltage in volts to ten decimals. If the sensor does not answer, the sketch prints ERROR: Can't init the sensor! Check connections! and stops.

Functions used​

ElectrochemicalGasSensor(sensorType _t, uint8_t _adcAddr = DEFAULT_ADC_ADDR, int _configPin = -1)returns none

Creates the sensor object. The sensor type holds the front-end and conversion settings for one gas cell. The address decides how begin() talks to the board: an address from 0x30 to 0x37 goes through the board's ATtiny404; any other address, such as the default 0x49, goes to an ADS1115 at that address and to the LMP91000 directly.

ReturnsConstructor.

Parameters

NameTypeDescription
_tsensorTypePreset from the library (SENSOR_CO, SENSOR_NO2, SENSOR_SO2, SENSOR_O3, SENSOR_NO, SENSOR_H2S, SENSOR_NH3, SENSOR_CL2) or a custom sensorType. For this board's cell, pass a custom one with 460 nA per ppm.
_adcAddruint8_tThe address begin() talks to. This example passes 0x49.
_configPinintA GPIO the library holds high and pulls low while it configures the LMP91000 directly. Ignored at addresses from 0x30 to 0x37. This example passes 32.
bool begin()returns bool

Starts I2C and picks how to talk to the board from the address. With an address from 0x30 to 0x37 it checks that the board's ATtiny404 answers and sends it the ADC gain from the sensor type with the slowest data rate; each command waits up to 500 ms for an answer. With any other address it starts an ADS1115 at that address with the same gain and data rate, sets the configuration pin as an output held high, and pulls it low while it configures the LMP91000. Call it once before reading.

Returnstrue when every step succeeded, false otherwise.

void setCustomZeroCalibration(double calibration)returns void

Sets the zero-point offset: a voltage that getPPM() adds after subtracting the front end's internal zero and before converting to PPM. It replaces the sensor type's internal zero calibration for this object.

ReturnsNothing.

Parameters

NameTypeDescription
calibrationdoubleOffset in volts. The example passes customCalibration (0.05).
double getPPM()returns double

Reads the ADC, subtracts the front end's internal zero (2.5 V times its percentage), adds the zero calibration, then divides by the amplifier gain and by the cell sensitivity in nA per ppm from the sensor type. A result below 0 is returned as 0. With ELECTROCHEMICAL_SENSOR_DEBUG defined it also prints the three voltages.

ReturnsGas concentration in PPM.

Putting it together​

Create the sensor, call begin(), then call setCustomZeroCalibration() with your offset before you take readings. Run the sketch where there is none of the target gas, change customCalibration, and upload it again until the voltage after calibration is roughly 0.