Start with the symptom you can observe. Apply one change at a time, then retry the smallest Quick Start example.

SymptomCheckFix
ERROR: Can't init the sensor! Check connections and jumper address!begin() returns false when the board's ATtiny404 does not answer at the address you passed to the constructor, or reports an error.Reseat the Qwiic cable at both ends, or check 3V3, GND, SDA1 and SCL1 on the header. Make sure the address in your code (0x30 in the bridgeBoard example) matches the ADDRESS switch.
The reading stays at 0The libraries return 0 whenever the calculated concentration is negative. The examples also use a preset for another gas, not this oxygen cell.Write your own sensor configuration for the SGX-4OX-ROHS cell (−600 mV bias), as the Arduino and MicroPython example pages describe.
Readings drift right after power-upAfter power-up the cell can take up to 15 minutes to settle.Leave the board powered and wait before you trust a reading.
Two boards on one bus do not both answerEvery board starts at 0x30.Give each board its own address from 0x30 to 0x37 with its ADDRESS switch, and create one sensor object per address.
The board was connected to 5 VThe board is 3.3 V only and has no regulator.Power it from 3.3 V only.

Still blocked?​

Ask on the community forum. Include the following so someone can reproduce it:

  • product name and SKU: Soldered Electrochemical O2 Gas Sensor Breakout (Qwiic), 333394
  • controller board and software environment
  • library or module revision
  • complete error text or unexpected output
  • connection method and power source
  • a clear photo of the wiring